Master thesis (2026) - T.J. Jeletich, J. Dankelman, R.M. Oosting, D. Dodou
Laparoscopic surgery significantly improves clinical outcomes over open procedures by reducing infection risks, postoperative pain, and recovery times. However, in low- and middle-income countries (LMICs), the widespread adoption of minimally invasive surgery remains severely restricted by the high acquisition costs, continuous electrical power demands, and fragile optical architectures of conventional laparoscopic tower systems. This thesis presents the design, development, and validation of an ultra-low-cost, modular USB laparoscope engineered as a direct replacement for commercial instruments in routine surgical procedures.

Adopting a human-centered, context-driven design methodology, functional requirements were formulated through a literature review, live surgical observations, and semi-structured interviews with healthcare professionals experienced in low-resource environments. These evaluations established that device robustness, ease of reprocessing, and on-site repairability decisively outweigh kinematic versatility and minor cost differences.
In the preliminary phase, four commercial-off-the-shelf (COTS) CMOS camera modules were quantitatively benchmarked against state-of-the-art clinical laparoscopic towers (Stryker 1688 and Olympus OTV-S7) using standardized ISO optical metrics. While the commercial reference systems demonstrated wider diagonal fields of view and superior low-light robustness, the low-cost CMOS modules exhibited lower color reproduction errors (Delta E and Delta C) and significantly reduced geometric barrel distortion (under -1.8% versus up to -4.9%). The 2-megapixel GC2755 module was selected for prototype integration due to its superior spatial sharpness at functional working distances (20–50 mm) and its compact footprint, which uniquely enabled a fixed 30-degree oblique viewing angle within a standard 12 mm outer diameter stainless steel shaft.
A functional prototype was fabricated incorporating an in-shaft distal housing with isolated forward-facing LED channels, an independent acrylic optical window, and a modular, two-part PETG handle featuring an elastomeric O-ring seal, brass cable gland, and tactile orientation indicator. Manufactured at a total component bill of materials (BOM) cost of €45.53, the scope interfaces directly via a single 5V USB cable to a standard laptop, eliminating infrastructural dependencies on external light sources and dedicated processing units.

Mechanical characterization confirmed structural resilience under operational tensile (25 N), torsional (1.7 Nm), and transverse bending loads. Thermal testing verified that steady-state operating temperatures at the distal tip plateau at 38.1°C, remaining safely below the 48°C tissue-injury threshold while closely matching physiological core body temperature (37°C), providing an intrinsic anti-fogging effect. Finally, an evaluative usability study with six practicing surgical professionals yielded a median System Usability Scale (SUS) score of 83.75 (mean 83.75 ± 9.05), substantially exceeding the 68-point clinical acceptability baseline. The participating surgeons confirmed that the prototype can be operated intuitively without retraining and possesses adequate optical fidelity for diagnostic and basic surgical procedures, establishing a viable foundation for accessible minimally invasive surgery in resource-constrained settings. ...

A multi-factor validation of vela™ vacuum extractor

Master thesis (2026) - A.M. Memelink, R.M. Oosting, J. Dankelman

Background: To reduce the huge amount of waste in healthcare, hospitals must switch to reusable medical devices. While single-use devices, such as the Kiwi OmniCup®, are currently standard in many clinical settings, they contribute to medical waste. To address these challenges, LAYCO has developed vela™, a reusable vacuum extractor designed to provide a sustainable alternative. However, unlike disposables, a reusable device operates within a continuous lifecycle. Its safe and effective use is therefore entirely dependent on two distinct groups: the logistical Central Sterile Services Department (CSSD) users who reprocess the device, and the clinical users who operate it. Objective: This study aimed to develop an Implementation Blueprint to ensure the correct clinical introduction of the reusable vela™ vacuum extractor by establishing a clinical baseline, validating user training across both user groups, and quantifying the influence of procedural factors and clinical use errors on the applied traction force. Methods: This research utilized a multi-phase approach. First, a retrospective cohort study established a clinical baseline of manual vacuum extractions at the Leiden University Medical Center (LUMC). Second, the CSSD preparation was validated through a randomized (dis)assembly study among Central Sterile Services Department students, evaluating use errors. Third, a randomized phantom study evaluated clinical use errors among medical interns, comparing standard Instructions for Use (IFU) against video-based training. Finally, a technical evaluation utilized an inline load-cell to objectively quantify the effect of procedural variations and clinical use errors on applied traction forces. Results: The clinical baseline revealed a procedure success rate of 97.0% and a severe perineal trauma (OASI) incidence of 2.0%. The CSSD and clinical evaluations indicated a descriptive trend toward fewer use errors with video-based training compared to instructions alone, while both groups demonstrated a statistically significant learning effect over time. Technical evaluations showed that peak traction forces (<130 N) remained safely below established clinical risk thresholds and indicated that peak traction forces were operator-dependent. Conclusion: This study established a localized clinical baseline and demonstrated that targeted video training effectively supports novice users in the device-specific handling of the vela™ vacuum extractor, as evidenced by a descriptive trend toward reduced use errors and a statistically significant learning effect. This Implementation Blueprint provides a comprehensive benchmark for the effective and sustainable clinical introduction of the vela™ device. ...

Visual prostheses aim to restore functional visual perception in individuals who are blind. Phosphene mapping is essential for these devices, as it establishes the relationship between implanted electrodes and locations in the visual field. Current approaches rely heavily on patient feedback, making the mapping process time-consuming and difficult to scale to high-density electrode arrays.

This thesis investigates Subthreshold Electrode Mapping (SEM) as a framework for automatically reconstructing the relative geometry of an implanted electrode array using subthreshold electrical stimulation. In SEM, stimulation signals recorded across the array are used to estimate inter-electrode distances, which are subsequently used to reconstruct the relative electrode positions. These positions can then be combined with a retinotopic model to support the phosphene mapping process.

The performance of different amplitude estimation approaches was investigated using analytical derivations and Monte Carlo simulations under several synthetic noise conditions. A frequency-domain approach consistently provided the most accurate amplitude estimates under the simulated conditions. The resulting distance estimates were incorporated into a precision-weighted geometry reconstruction framework based on multidimensional scaling and weighted SMACOF. Simulation results demonstrated that SEM can reconstruct the relative electrode geometry with high accuracy and that reconstruction performance improves with increasing stimulation budget.

Overall, the results demonstrate the feasibility of SEM as a potential approach for relative electrode localization and provide a simulation framework for evaluating its performance. Further experimental validation is required to assess its performance under realistic neural tissue and noise conditions. ...

A proof-of-concept validation in a cervicovaginal phantom model with gynaecologists

Master thesis (2026) - W.M. Oosterhof, J. Dankelman, Marcus Rijken, Nick J.van de Berg, P. Breedveld
Cervical cancer represents an important global health challenge, particularly in low- and middle-income countries (LMICs), where access to screening, treatment, and specialised equipment may be limited. Cold knife conisation (CKC) can play an important role in the treatment and diagnosis of cervical disease, but the procedure requires controlled excision to obtain a complete cone-shaped specimen. This procedure can be technically demanding because CKC is affected by limited visibility, bleeding, anatomical variability, and the need to perform an angled excision while guiding the instrument accurately. A first version of the conisator was previously developed to support excision movements during CKC, but still required manual stabilisation of the bearing system. Therefore, the need for an integrated fixation mechanism formed the central design challenge addressed in this research.

A problem analysis was performed using literature research, input from gynaecologists, and evaluation of the first conisator prototype. Based on this analysis, a balloon-based fixation concept was selected and developed into a modular proof-of-concept prototype. The final design consisted of a manually inflated balloon retained within a bearing system, allowing temporary fixation around the cervix while the excision components could rotate. A phantom model was developed to simulate the cervicovaginal surgical field. The prototype was validated by eight gynaecologists through a simulated CKC procedure and questionnaire-based evaluation of task performance, design criteria, and requirements. In addition, a specimen measurement test was performed to assess the dimensional performance of the excised cone-shaped specimens.

The design criteria were generally evaluated positively, with an overall weighted score of 3.7 out of 5. Most procedural steps, including balloon inflation, stabiliser placement, and device removal, could be performed easily. The main design criteria requiring further optimisation were device control, cervix movement, cervical visibility, and user confidence. Cervical visibility through the inner ring remained limited, which made it more difficult to observe the
excision and locate the cervical canal during insertion. Overall, 11 of the 13 requirements were fulfilled. The two requirements that were not completely fulfilled concerned rotational resistance and anatomical fit. Rotational resistance received 63% positive responses, while anatomical fit was not fully achieved because the width of the fixation mechanism exceeded the reported mean minimum fornix width. The specimen measurement test showed that the conisator approximated the preset specimen width of 24 mm, with measured widths ranging from 21 to 26~mm and a mean width of 23.1 mm. Although minor variation remained between measurements, the specimen widths stayed close to the preset target. This indicates that the conisator effectively guided the excision towards the intended width, despite sensitivity to alignment, device orientation, blade sharpness, and scalpel-holder flexibility. From an LMIC perspective, the design fulfilled several key functions through electricity-independent and tool-free operation, modular construction, cost efficiency, design for sterilisation, and ease of use.

Overall, this research demonstrated that an integrated balloon-based fixation mechanism can provide temporary fixation and support controlled use of the conisator during simulated CKC. By integrating fixation into the instrument, this work addresses the design gap of stabilising the conisator at depth within the cervicovaginal space. The final proof-of-concept prototype therefore provides a promising basis for further development towards a reusable instrument for supporting CKC, particularly in LMIC settings. ...
Master thesis (2026) - M.S. Sundaram, J. Dankelman
Flexible endoscopy is an essential diagnostic and therapeutic tool, yet its high capital cost, complex maintenance requirements, and monolithic architecture render it largely inaccessible in low-resource settings. When a single mechanical subsystem fails in a standard commercial endoscope, the entire unit is frequently retired due to the lack of local repair infrastructure. This research presents the design, fabrication, and validation of a low-cost, modular, flexible colonoscope engineered to overcome these barriers. The prototype utilizes a magnetic quick-swap cassette architecture that completely decouples the tensioned steering cables and the distal bending section from the primary ergonomic handle. This modularity enables rapid field replacement of mechanical components without requiring specialized tools or device disposal.

Fabricated entirely from commercial off-the-shelf components and additive manufacturing, the device features active four-way mechanical steering, a continuous working channel for surgical instrument passage, and integrated fluid pathways for suction, irrigation, and insufflation. During benchtop validation, the continuum bending section achieved up to 214° of articulation in both the upward and downward directions while fully sheathed, thereby exceeding standard clinical thresholds. The fluidic subsystems were independently verified to operate without physical fluid leakage. Additionally, a post-procedure purging routine, controlled by an integrated microcontroller, was successfully implemented to ensure fluid clearance. The total Bill of Materials (BOM) for the functional system was constrained to ≈ €128. This work demonstrates that a fully therapeutic modular endoscope can be produced at a fraction of the cost of commercial systems, providing a foundational platform for improved healthcare accessibility in low- and middle-income countries. ...

Two phantoms to train for episiotomy and its repair

Master thesis (2026) - J.J.B. Smit, J. Dankelman, D. Khalid Hassan Abubakr, R.M. Oosting, Evelien M. Sandberg, Irene Schouten-Koning
Mediolateral episiotomy is performed to reduce the risk of severe perineal tearing during childbirth. Although the procedure is used less frequently than in the past, practitioners still need to perform it correctly when indicated. Current training methods mainly rely on animal tissue, simple low-fidelity models, or expensive commercial simulators. Most existing models do not provide realistic anatomy and tactile feedback or allow training in both episiotomy placement and repair.
This thesis aimed to develop a low-cost, high-fidelity episiotomy trainer that allows practitioners to train both mediolateral episiotomy placement and repair. A literature review and interviews with three gynaecologists and a surgeon were conducted to determine the learning goals, design criteria, and requirements for the phantom. Based on these requirements, five concepts were developed and compared using a Harris profile. The selected concept consisted of two separate silicone phantoms: one for episiotomy placement and one for episiotomy repair. Several prototype iterations were manufactured and evaluated to improve the anatomy, material properties, and manufacturability of the models. 12 gynaecological and obstetric practitioners tested the prototypes during evaluations at a Dutch hospital and a perineal laceration workshop.
The final episiotomy placement phantom allows training of the correct finger position, scissor placement, and incision angle during crowning of the foetal head. The repair phantom includes the skin, vaginal wall, subcutaneous tissue, bulbospongiosus muscle, transverse perineal muscle, anal sphincter, and rectum, allowing participants to practise all essential steps of episiotomy repair. Both models met the predefined learning goals and were designed for local manufacturing at low cost.
The two developed phantoms form a realistic and affordable method for training mediolateral episiotomy placement and its repair. They offer an alternative to existing low-fidelity and high-cost commercial models and can contribute to improving practical training in obstetrics.
...
Master thesis (2026) - R. Brands, D. Khalid Hassan Abubakr, J. Dankelman, R.M. Oosting, Evelien M. Sandberg, Nick van de Berg
Caesarean section (CS) can be an essential surgical procedure, still maternal and perinatal outcomes after CS remain substantially worse in many low- and middle-income countries (LMICs) as compared to high-income countries. Simulation may complement apprenticeship-based training as it allows trainees to practise procedural steps in a safe environment. However, many commercial CS simulators are too expensive for a low resource setting, and existing low-cost trainers often only address isolated tasks. This project developed a low-cost, modular CS training phantom for practising the procedural sequence from skin incision to uterine closure.

The design builds on the previous Salama CS phantom and adds uterine incision, blunt extension, fetal and placental extraction, and an impacted fetal head scenario. Design requirements and learning goals were set up based on a clinical and design context analysis, and concepts were systematically generated, selected, and improved through iterative prototyping. Selective fidelity was applied: low-cost textiles and foam represented the replaceable abdominal layers, whereas a reinforced silicone strip was used at the lower uterine segment for incision, blunt extension, fetal extraction, and suturing.

Four clinical experts evaluated an initial prototype using a five-point Likert-scale questionnaire and open questions which addressed the ability to practise individual procedural steps, anatomical realism, and educational value. Mean scores ranged from 3.7 to 5.0 for procedural practice, from 4.0 to 5.0 for anatomical realism, and from 4.0 to 5.0 for educational value. Qualitative and quantitative feedback was used for the design of the final prototype which was inspected by two experts who provided qualitative feedback. The final prototype had an estimated one-time material cost of €19.21, which was 80% lower than the €94.02 Salama benchmark, and an estimated recurring material cost of €4.23 per use. These values exclude the externally supplied fetal model and materials obtained from existing workshop stock.

These findings demonstrate the potential of a low-cost, modular training phantom that includes all the major CS steps and an impacted fetal head scenario in supporting safe CS practice. Evidence of learning effectiveness, durability, local manufacturability, and suitability for LMIC implementation requires further validation of the final prototype with the intended trainee population. ...
Master thesis (2026) - N.M. Jonk, R. Straathof, J. Dankelman, Nick J.van de Berg
Purpose. The ARCHITECT applicator is a 3D-printed, patient-tailored intracavitary/interstitial (IC/IS) brachytherapy (BT) applicator concept in which channel configurations are automatically generated and optimised. Previous work showed that dosimetric advantages of this applicator are not uniform across patients, and that cases with vaginal tumour extension may be more challenging to treat. This study investigated the interaction between ARCHITECT applicator configuration generation and automated treatment planning in BiCycle, with the aim of identifying potential workflow adaptations to improve treatment plans for patients with vaginal extension.
Methods. Treatment plans of the first BT fraction were analysed for 85 patients with LACC previously treated with the Utrecht or Venezia applicator. Of these patients, 32 had vaginal extension greater than 10 mm. In this study, three objectives were addressed. First, dosimetric analysis compared treatment plans generated for the conventional applicator configuration used clinically with plans generated for the ARCHITECT applicator. Second, the interplay between the ARCHITECT RTPlan conversion and automated planning software BiCycle was assessed using a MATLAB-based validation tool. This tool systematically modified the applicator setup in the RTPlan to evaluate the effect on BiCycle-generated treatment plans and validate BiCycle’s handling of applicator configuration. Third, three strategies were investigated to improve ARCHITECT treatment plans for a difficult to irradiate patient case with vaginal extension: adjustment of needle dwell time smoothing, inclusion of inferior dwell positions during geometric coverage planning, and generation of dedicated IC channels.
Results. The ARCHITECT applicator showed dosimetric advantages over the conventional applicator in both patients with and without vaginal extension. In patients without vaginal extension, this advantage was mainly reflected in improved OAR sparing. In patients with vaginal extension, the advantage was more apparent in target coverage. The RTPlan processing validation showed that systematic modifications of the applicator setup resulted in expected changes in BiCycle-generated treatment plans. The strategy involving the addition of dedicated IC channels showed the most favourable effect on ARCHITECT plan quality for vaginal extension.
Conclusion. This study indicates that the ARCHITECT applicator can be used for patients with vaginal extension, although the conventional applicator may remain preferable in specific patient cases. The RTPlan processing validation supported quality assurance by confirming that BiCycle responded as expected to systematic changes in applicator setup. Including IC needle channels during coverage planning in the ARCHITECT workflow was the most promising strategy to improve treatment plans for patients with vaginal extension. This strategy should be further refined and evaluated for a larger patient cohort with vaginal extension to assess its future potential. ...
Total hip arthroplasty (THA) is one of the most frequently performed orthopedic surgeries. Due to indications such as chronic infection and femoral stem loosening, revision total hip arthroplasty (rTHA) is sometimes required. However, as uncemented prostheses are designed to fixate due to bone ingrowth, stem removal introduces complex surgical challenges. Current extraction instruments do not conform to the complex curvature of the medial surface of the stem. In addition, following the lateral surface along the entire stem length and reaching the distal region remains challenging. This research focuses on the medial surface, optimizing the balance between conforming to the complex medial curvature and maintaining efficient axial impact transmission. Three different medial osteotome prototypes were designed and manufactured, varying in material type and geometry. A three-point bending test was conducted to quantify how material and geometry influence flexural compliance and clinical safety of the prototypes. Furthermore, a Rockwell C hardness test further was performed to evaluate the hardness of the osteotomes and a sharpness test was conducted to identify the cutting edges. Finally, the final osteotome insertion and implant loosening test reflects the prototypes' performance in a simulation of the clinical setting, combining conforming to the medial curve of uncemented femoral stems while cutting through a polyurethane foam bone model. The results suggest that introducing enhanced flexibility does not necessarily compromise axial impact transmission, offering a promising solution for medial osteotomes used in total hip revision surgery. However, further research should demonstrate whether this introduced flexibility enhances medial surface loosening of different types of prostheses in human femoral trabecular bone and effectively decreases the risk of calcar fracture. ...
Master thesis (2026) - P.M.J. van der Hout, J. Dankelman, N.A. White, Koen Van der Bogt, T. Horeman
Purpose – Patients with end-stage kidney disease often rely on haemodialysis, which requires a surgical connection between a vein and an artery. However, the resulting persistent, high and turbulent blood flow leads to vascular and cardiac complications. Recently, a valve mechanism was developed to regulate the anastomosis by manipulating a synthetic graft. However, it is unknown how the valve mechanism will function in vivo. Improper compression can lead to low or stagnant blood flow, which is undesirable due to the increased risk of thrombus formation. Therefore, this thesis investigates an alternative valve mechanism that does not rely on graft manipulation, aiming to improve resistance to thrombus formation.
Objective – The research goal of this thesis is to develop and validate a proof-of-concept of a valve mechanism for a dynamic artery–vein connection based on ananastomotic shape-preserving working principle.
Method – This research is conducted according to a systematic roadmap that describes the process of translating biomedical research to medical devices. The creative process within this roadmap is based on the Delft Design Approach. Generated concepts are scored using a Harris profile based on criteria minimising health risks to patients. The working principle is validated through friction force, leakage, and lifecycle experiments. The hardness of the seals affects the resulting friction force. To characterise its range, three seal configurations are evaluated: soft (20A), medium (40A), and hard (60A).
Results – The outcome of this thesis is a shape-preserving valve mechanism that regulates the anastomosis. Regulation is achieved via a one-degree-of-freedom linear system that extends and retracts a Nitinol blade between two polyurethane seals. Preload on the seals is achieved by compressing the side plates against the housing. The experiments show that an increase in seal hardness results in a higher friction force. Furthermore, the risk of buckling increases with increasing hardness. Moreover, leakage occurred at the lowest seal hardness configuration. The 40A seal configuration outperformed the others and demonstrated consistent performance across all experiments.
Conclusion – From this research, the conclusion is drawn that a valve mechanism with an anastomotic shape-preserving working principle shows potential for further development. Based on the experiments, the 40A hardness seal configuration is deemed to have the most potential. Future studies should focus on feasibility tests and mechanical development. Then, mechanical optimisation and medical development should focus on how the mechanism functions in vivo, to investigate the operating principle, leakage risk and how the device will behave when subjected to the foreign body response.
...
Master thesis (2026) - T. Vermeijden, J. Dankelman, Nick van de Berg, Laurie van de Weerd, Mark Witteveen, Benno Hendriks
Accurate intraoperative assessment of tumour margins remains a challenge in the surgical treatment of vulvar cancer. Incomplete resection increases the risk of recurrence while excessive tissue removal negatively affects the patient's QoL. HSI offers a non-invasive technique that captures both spatial and spectral information, enabling tissue classification based on spectral characteristics. However, clinical applicability is limited by low spatial quality, highlighting the need for techniques such as super-resolution reconstruction that can improve spatial quality. This study investigates the potential of HSI combined with deep learning and super-resolution reconstruction for tumour margin assessment in vulvar cancer.

A reproducible pipeline was developed for data acquisition from patients, data preprocessing, super-resolution reconstruction and classification using a 3D CNN. The acquisition process included both ex vivo and in vivo imaging during surgery. Corresponding RGB images and hypercubes were acquired and aligned for ROI annotation. Preprocessing consisted of calibration and correction steps to improve the comparability of spectral signatures, while super-resolution reconstruction was used to enhance the spatial quality of the hypercubes. The approach was evaluated on both ex vivo and in vivo data to assess the model's ability to distinguish between tumour and non-tumour tissue under different training and testing configurations. In particular, the generalisability of models trained on ex vivo data to in vivo data was evaluated.

The results demonstrate that HSI-based models achieved strong classification performance for tumour classification. Furthermore, models trained on ex vivo data showed the ability to generalise to in vivo data. Super-resolution reconstruction enhanced the spatial quality of hypercubes without negatively affecting the classification performance. This enables visual interpretation of the hypercubes and supports clinical implementation.

In conclusion, this thesis demonstrates that the combination of hyperspectral imaging, super-resolution reconstruction and deep learning is a promising approach for tumour margin assessment in vulvar cancer. The proposed pipeline provides a solid foundation for future research and supports the continued development of real-time, intraoperative tumour margin assessment using HSI. Limitations of dataset size and balance, preprocessing steps and computational demands restrict clinical implementation currently. These challenges highlight the need for further research. ...
Master thesis (2026) - S.A. Haupt, T. Horeman, J. Dankelman
Lower urinary tract dysfunction (LUTD) refers to a range of disorders affecting the lower urinary tract and impacts millions of individuals worldwide, diminishing quality of life. Urodynamic studies (UDS), including cystometry, are used to diagnose LUTDs by measuring intravesical bladder pressure to assess bladder function. Conventional cystometry relies on a dual-catheter configuration, in which a second abdominal catheter serves as a reference signal to distinguish motion artifacts from true detrusor contractions. While effective, this approach increases invasiveness, is costly, and may limit bladder monitoring under natural conditions. Ambulatory cystometry enables real-world symptom replication during daily activities, but still relies on dual catheters.

This thesis investigates two strategies to enable motion context awareness in single-catheter ambulatory cystometry, aiming to reduce invasiveness, improve accessibility, and maintain diagnostic reliability.

Part A explores a machine learning–based approach to distinguish motion artifacts from detrusor contractions using a single pressure signal. A structured, posture-aware signal processing and classification pipeline was developed and evaluated using cystometry data from ten patients. Multiple modelling strategies were evaluated, including feature-based classifiers, neural network architectures, and hybrid strategies. Neural networks demonstrated superior overall performance, although increasing architectural complexity did not result in noticeable performance gains. While results are limited by cohort size, the findings indicate that motion artifacts and detrusor events exhibit distinguishable pressure characteristics and establish a scalable analytical framework for future dataset expansion.

Part B addresses motion context awareness through hardware integration. The UroMonitor, a low-cost, single-catheter ambulatory cystometry prototype, was independently designed and developed. The system integrates a pressure sensor, inertial measurement unit (IMU), microcontroller, wireless connectivity, and on-board storage within a portable, battery-powered platform. By simultaneously acquiring pressure and motion signals, the device enables correlation-based identification and suppression of motion artifacts without requiring a second catheter. Phantom testing demonstrated reliable motion detection, with acceleration magnitude signals achieving a mean recall of 95.4\%. The prototype satisfied predefined physical, electrical, economic, physiological, and usability design constraints.

Together, these contributions demonstrate the technical feasibility of a low-cost, motion-aware, single-catheter ambulatory cystometry system through both machine learning and hardware-based strategies, providing a foundation for future clinical validation and scalable deployment, including in low-resource healthcare settings. ...

A Possible Solution to Reducing Saddle Discomfort

Master thesis (2026) - L.M. Lingmont, J. Dankelman, Nick J. van de Berg, P. Vink
Cycling plays a major role in daily transportation in the Netherlands and offers substantial health, economic, and environmental benefits. However, for individuals with vulvar conditions such as lichen sclerosus and lichen planus, cycling can cause significant discomfort or even become impossible. Existing saddle solutions do not adequately address the needs of all users, highlighting the need for a more personalised and adaptable approach.
The aim of this thesis was to design an adjustable bicycle saddle that enables localised pressure redistribution, and to evaluate how different configurations influence pressure distribution and perceived comfort in women.

The design process followed a structured approach, starting with defining functional requirements and generating multiple concepts using a morphological chart. These concepts were evaluated using a Harris profile, leading to the selection of a threaded peak system with a foam top layer. The final prototype was based on an existing saddle and incorporated adjustable levelling feet. Rotating the levelling feet allowed local height adjustments to modify the pressure distribution.

To evaluate the prototype, a repeated measures study was conducted with 24 healthy female participants. Four saddle conditions were tested: the original saddle and three configurations of the adjustable saddle. Saddle pressure distribution was measured using a pressure mapping system, and perceived comfort was assessed through questionnaires. Participants were asked to provide comfort scores and to produce drawings representing their perceived saddle pressure for each tested saddle.

The results showed that lowering the levelling feet of the adjustable saddle reduced pressure in the targeted regions without increasing pressure in surrounding areas. However, these localised pressure changes were not visible in saddle pressure drawings made by the participants. The original saddle produced significantly higher pressure in the left and right regions and showed lower comfort scores compared to all configurations of the adjustable saddle. No significant differences in comfort score were found between the different adjustable configurations.

In conclusion, the developed adjustable saddle successfully enabled localised pressure reduction. Although these changes were measurable, they were not always perceptible to users, suggesting that larger or more pronounced adjustments may be required. However, the intended user population (e.g. individuals with saddle-related complaints) might benefit from the pressure relief regions. Future research should therefore focus on evaluating the design in the intended user population over extended periods of use. ...

Context-Driven Design of a Camera Actuator for Low-Resource Settings

This thesis presents the design, realisation, and evaluation of a robotic laparoscopic camera assistant intended to support training and practice in low-resource settings, compatible with gasless laparoscopic techniques. The work is grounded in a context-driven design approach, informed by structured end-user engagement.A compact actuator was developed to control pitch, yaw, and insertion of a laparoscope. The system was verified against functional and non-functional requirements through mechanical testing and operator validation. Results demonstrate that the actuator enables stable camera positioning across the required range of motion, remains fully backdrivable, and maintains image stability while using relative position control. Limitations were observed in achievable actuation speeds and load resistance, which were traced back to coupled design trade-offs between motor gearing, backdrivability, and friction in the insertion mechanism. Validation with clinicians showed that the actuator allows consistent establishment of the required view and feasible task execution, particularly in conventional laparoscopic configurations.The findings illustrate that meaningful camera assistance can be achieved using relatively simple mechanics and control architectures when design priorities are aligned with contextual constraints. While the current prototype is primarily suited for training and simulated use, the identified trade-offs and proposed design adaptations provide a clear pathway toward increased clinical implementation. More broadly, this work demonstrates how context-driven design can guide the development of surgical robotic systems that prioritise accessibility and practical impact over technological complexity. ...
Master thesis (2026) - L.J. Stakenburg, R. Straathof, J. Dankelman, N.J.P. van de Berg
Purpose: Large or non-superficial vaginal tumours or vaginal involvement require interstitial (IS) needles in brachytherapy. When such target volumes extend below the loading plane, the IS needles of ring and ovoid applicators cannot reach them, so freehand perineal needles are currently used. These demand considerable physician skill and involve long needle paths. This study developed an automated planning method for curved paravaginal needles through a cylindrical applicator and compared target coverage with freehand needles.

Methods: For nine patients treated with freehand needles (configFN), a virtual cylindrical applicator was constructed. From standardised entry points, 2,000 candidate single-curved needle trajectories were sampled under geometric, mechanical and OAR constraints. An integer program then selected the needle configuration maximising CTVIR coverage (configopt), which was compared with configFN in coverage and total needle path length.

Results: The candidate sets densely covered the feasible exit regions, with a median maximum attainable coverage of 98.7%. Median coverage increased from 55.1% for configFN to 79.4% for configopt (p = 0.027), improving in eight of nine patients. Median total needle path length decreased from 72.1 to 32.6 cm. The median total time needed for creating a complete plan was 774 seconds.

Conclusions: Automated planning of curved needles through a cylindrical applicator achieved higher geometric target coverage with shorter total needle length than freehand perineal needles, within a clinically acceptable computation time. The exception was tumours in the lower vagina, where limited space for needle curvature may still favour perineal needles. Future work should include dwell time optimisation and DVH-based dosimetric comparison, followed by manufacturing and clinical evaluation of the applicator.
...

Development and Experimental Validation of Bibble

Master thesis (2026) - J. van den Berg, J. Dankelman, R.M. Oosting, R. Breda
Access to respiratory support remains limited in low-resource hospitals, where equipment and maintenance costs, shortages of consumables, and limited staffing and monitoring capabilities constrain the use of conventional ventilators. This thesis aimed to develop and evaluate Bibble, a simple bilevel positive airway pressure (BiPAP) device for low-resource hospitals in Nepal.

An online questionnaire investigated respiratory care capabilities, clinical needs and practical constraints in Nepali hospitals. The findings were combined with clinical literature and safety considerations to establish a Programme of Requirements. A step-by-step design process was used to develop and evaluate concepts for pressure regulation, switching and supporting functions. Concepts were assessed for simplicity, repairability and their ability to deliver BiPAP. The selected concept was developed into a functional prototype using silicone tubing, a servo motor and fused deposition modelling. Bench tests characterised pressure delivery, and performance was compared with the BubbleVent and VN500.

The questionnaire identified a need for additional respiratory support capacity and highlighted the importance of simple operation, low cost, maintainability, compatibility with different gas infrastructures and efficient oxygen use. Bench testing demonstrated that the prototype could reliably alternate between two pressure levels within a clinically relevant range. It produced a smoother pressure waveform than the BubbleVent, although pressure rise and fall were slower than those of both comparison devices.

This thesis demonstrates the feasibility of manufacturing a simple bubble-based BiPAP prototype capable of alternating between two pressure levels under bench conditions. Further development should improve pressure-setting accuracy and reduce pressure transition times. These findings provide a foundation for continued development of respiratory support suited to low-resource hospitals and their practical constraints. Additional testing and validation are required to establish whether the device can meet the safety and performance requirements of its intended setting before conclusions can be drawn about its clinical safety and effectiveness in patient care. ...
Master thesis (2026) - F.E.M. Matter, J. Dankelman, Azhaar Ramsaroep, Daphne van Scheppingen, Benno Hendriks
Problem: Respiratory rate (RR) is a sensitive early indicator of clinical deterioration. However, on general wards it is still largely assessed through intermittent manual spot-checks. Camera-based monitoring offers a continuous, unobtrusive alternative, but research has predominantly focused on mean RR, even though breath-to-breath variability provides additional information on respiratory control. This study investigated whether a custom camera-based system can detect variability in inspiratory and expiratory time (Ti, Te), breathing rate (BRV), and breathing depth (BDV). Methods: An algorithm was developed that extracts a thoracoabdominal motion signal using Farneback dense optical flow, followed by adaptive peak and trough detection and breath-by-breath matching to a reference. The algorithm was first evaluated on a programmable respiratory phantom across nine combinations of breathing depth (1-21.8 mm) and rate (8-30 brpm), with a laser displacement sensor as reference. Subsequently, it was validated in 24 two-minute supine recordings from eight healthy adults, with capnography (for Ti, Te, BRV) and a respiratory inductance plethysmography (RIP) belt (for BDV) as references. Results: On the phantom,
Ti, BRV, and BDV showed high correlations with the reference (r=0.84–1.00), while Te showed more variable correlations despite comparable absolute errors. Agreement generally decreased with decreasing breathing depth. In human subjects, BRV showed the highest mean correlation with the reference (r=0.95) and a mean difference in coefficient of variation (CV) of 0.9%, followed by BDV (r=0.83; ∆CV=−3.1%). Ti and Te showed moderate agreement (r=0.75 each) with CV of 5.4% and −0.2%, respectively. Conclusions: A camera-based monitor using optical flow can detect breathing variabilities (Ti, Te, BRV, and BDV) under controlled conditions, with BRV showing the highest correlation in human validation. Agreement decreased with shallower breathing in the respiratory phantom. In human recordings, part of the observed disagreement could be attributed to artefacts in the reference signals. Further validation using longer recordings, varying body positions, and diverse patient populations is needed before clinical application. ...
Distal nozzles in multioutlet sterilization rods receive reduced pressure due to axial pressure losses, limiting cleaning capacity when outlets are plugged to maintain performance. This thesis develops and evaluates a passive, flow-powered rotary-valve concept (GreenClean) intended to generate low-frequency pulsating jets while improving outlet-to-outlet distribution within a half-DIN sterilization basket. A five-outlet prototype was designed and experimentally validated using time-resolved pressure and flow measurements, progressive outlet-blockage tests, and supporting CFD as an idealised leakage-free baseline.

The prototype produced stable pulsation at approximately one cycle per 62–69 s (14–16 mHz), well below the 2 Hz requirement, and pulsation persisted under progressive and full blockage. However, uniform distribution was not achieved experimentally: mean deviations reached ∼ 25% in outlet flow and ∼ 39% in outlet pressure. In contrast, sealed CFD predicted outlet total-pressure deviations within ±4.1%, indicating that the concept can approach the distribution target in principle. Additional tests revealed substantial nonideal behaviour, including a leakage fraction of ηleak ≈ 0.48 at the connection/ sealing region and a marked reduction in available head in a jet height comparison. A stall-torque estimate of τstall ≈ 0.28 Nm provides an order-of-magnitude design margin for future sealing and alignment improvements. Environmental resistance was not validated and requires dedicated material and seal selection for washer-disinfector conditions.
...