Camera-Based Detection of Breathing Variability: Algorithm Design and Validation

From Respiratory Phantom to Human Validation Study

Master Thesis (2026)
Author(s)

F.E.M. Matter (TU Delft - Mechanical Engineering)

Contributor(s)

J. Dankelman – Mentor (TU Delft - Mechanical Engineering)

Azhaar Ramsaroep – Mentor

Daphne van Scheppingen – Mentor

Benno Hendriks – Graduation committee member (TU Delft - Mechanical Engineering)

Faculty
Mechanical Engineering
More Info
expand_more
Publication Year
2026
Language
English
Graduation Date
02-10-2026
Awarding Institution
Delft University of Technology
Programme
Biomedical Engineering, Medical Devices
Faculty
Mechanical Engineering
Page Views
48
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Problem: Respiratory rate (RR) is a sensitive early indicator of clinical deterioration. However, on general wards it is still largely assessed through intermittent manual spot-checks. Camera-based monitoring offers a continuous, unobtrusive alternative, but research has predominantly focused on mean RR, even though breath-to-breath variability provides additional information on respiratory control. This study investigated whether a custom camera-based system can detect variability in inspiratory and expiratory time (Ti, Te), breathing rate (BRV), and breathing depth (BDV). Methods: An algorithm was developed that extracts a thoracoabdominal motion signal using Farneback dense optical flow, followed by adaptive peak and trough detection and breath-by-breath matching to a reference. The algorithm was first evaluated on a programmable respiratory phantom across nine combinations of breathing depth (1-21.8 mm) and rate (8-30 brpm), with a laser displacement sensor as reference. Subsequently, it was validated in 24 two-minute supine recordings from eight healthy adults, with capnography (for Ti, Te, BRV) and a respiratory inductance plethysmography (RIP) belt (for BDV) as references. Results: On the phantom,
Ti, BRV, and BDV showed high correlations with the reference (r=0.84–1.00), while Te showed more variable correlations despite comparable absolute errors. Agreement generally decreased with decreasing breathing depth. In human subjects, BRV showed the highest mean correlation with the reference (r=0.95) and a mean difference in coefficient of variation (CV) of 0.9%, followed by BDV (r=0.83; ∆CV=−3.1%). Ti and Te showed moderate agreement (r=0.75 each) with CV of 5.4% and −0.2%, respectively. Conclusions: A camera-based monitor using optical flow can detect breathing variabilities (Ti, Te, BRV, and BDV) under controlled conditions, with BRV showing the highest correlation in human validation. Agreement decreased with shallower breathing in the respiratory phantom. In human recordings, part of the observed disagreement could be attributed to artefacts in the reference signals. Further validation using longer recordings, varying body positions, and diverse patient populations is needed before clinical application.