A multi-factor validation of vela™ vacuum extractor

Master thesis (2026) - A.M. Memelink, R.M. Oosting, J. Dankelman

Background: To reduce the huge amount of waste in healthcare, hospitals must switch to reusable medical devices. While single-use devices, such as the Kiwi OmniCup®, are currently standard in many clinical settings, they contribute to medical waste. To address these challenges, LAYCO has developed vela™, a reusable vacuum extractor designed to provide a sustainable alternative. However, unlike disposables, a reusable device operates within a continuous lifecycle. Its safe and effective use is therefore entirely dependent on two distinct groups: the logistical Central Sterile Services Department (CSSD) users who reprocess the device, and the clinical users who operate it. Objective: This study aimed to develop an Implementation Blueprint to ensure the correct clinical introduction of the reusable vela™ vacuum extractor by establishing a clinical baseline, validating user training across both user groups, and quantifying the influence of procedural factors and clinical use errors on the applied traction force. Methods: This research utilized a multi-phase approach. First, a retrospective cohort study established a clinical baseline of manual vacuum extractions at the Leiden University Medical Center (LUMC). Second, the CSSD preparation was validated through a randomized (dis)assembly study among Central Sterile Services Department students, evaluating use errors. Third, a randomized phantom study evaluated clinical use errors among medical interns, comparing standard Instructions for Use (IFU) against video-based training. Finally, a technical evaluation utilized an inline load-cell to objectively quantify the effect of procedural variations and clinical use errors on applied traction forces. Results: The clinical baseline revealed a procedure success rate of 97.0% and a severe perineal trauma (OASI) incidence of 2.0%. The CSSD and clinical evaluations indicated a descriptive trend toward fewer use errors with video-based training compared to instructions alone, while both groups demonstrated a statistically significant learning effect over time. Technical evaluations showed that peak traction forces (<130 N) remained safely below established clinical risk thresholds and indicated that peak traction forces were operator-dependent. Conclusion: This study established a localized clinical baseline and demonstrated that targeted video training effectively supports novice users in the device-specific handling of the vela™ vacuum extractor, as evidenced by a descriptive trend toward reduced use errors and a statistically significant learning effect. This Implementation Blueprint provides a comprehensive benchmark for the effective and sustainable clinical introduction of the vela™ device. ...