Development of a Low-Cost Caesarean Section Training Phantom for Use in Low- and Middle-Income Countries

Master Thesis (2026)
Author(s)

R. Brands (TU Delft - Mechanical Engineering)

Contributor(s)

D. Khalid Hassan Abubakr – Mentor (TU Delft - Mechanical Engineering)

J. Dankelman – Mentor (TU Delft - Mechanical Engineering)

R.M. Oosting – Mentor (TU Delft - Mechanical Engineering)

Evelien M. Sandberg – Mentor (Leiden University Medical Center)

Nick van de Berg – Graduation committee member (Erasmus MC)

Faculty
Mechanical Engineering
More Info
expand_more
Publication Year
2026
Language
English
Graduation Date
13-08-2026
Awarding Institution
Delft University of Technology
Programme
Mechanical Engineering, BioMechanical Design
Faculty
Mechanical Engineering
Page Views
90
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Caesarean section (CS) can be an essential surgical procedure, still maternal and perinatal outcomes after CS remain substantially worse in many low- and middle-income countries (LMICs) as compared to high-income countries. Simulation may complement apprenticeship-based training as it allows trainees to practise procedural steps in a safe environment. However, many commercial CS simulators are too expensive for a low resource setting, and existing low-cost trainers often only address isolated tasks. This project developed a low-cost, modular CS training phantom for practising the procedural sequence from skin incision to uterine closure.

The design builds on the previous Salama CS phantom and adds uterine incision, blunt extension, fetal and placental extraction, and an impacted fetal head scenario. Design requirements and learning goals were set up based on a clinical and design context analysis, and concepts were systematically generated, selected, and improved through iterative prototyping. Selective fidelity was applied: low-cost textiles and foam represented the replaceable abdominal layers, whereas a reinforced silicone strip was used at the lower uterine segment for incision, blunt extension, fetal extraction, and suturing.

Four clinical experts evaluated an initial prototype using a five-point Likert-scale questionnaire and open questions which addressed the ability to practise individual procedural steps, anatomical realism, and educational value. Mean scores ranged from 3.7 to 5.0 for procedural practice, from 4.0 to 5.0 for anatomical realism, and from 4.0 to 5.0 for educational value. Qualitative and quantitative feedback was used for the design of the final prototype which was inspected by two experts who provided qualitative feedback. The final prototype had an estimated one-time material cost of €19.21, which was 80% lower than the €94.02 Salama benchmark, and an estimated recurring material cost of €4.23 per use. These values exclude the externally supplied fetal model and materials obtained from existing workshop stock.

These findings demonstrate the potential of a low-cost, modular training phantom that includes all the major CS steps and an impacted fetal head scenario in supporting safe CS practice. Evidence of learning effectiveness, durability, local manufacturability, and suitability for LMIC implementation requires further validation of the final prototype with the intended trainee population.