Development of a balloon-based fixation mechanism for a cervical conisation instrument
A proof-of-concept validation in a cervicovaginal phantom model with gynaecologists
Master thesis
(2026)
-
W.M. Oosterhof, J. Dankelman, Marcus Rijken, Nick J.van de Berg, P. Breedveld
Cervical cancer represents an important global health challenge, particularly in low- and middle-income countries (LMICs), where access to screening, treatment, and specialised equipment may be limited. Cold knife conisation (CKC) can play an important role in the treatment and diagnosis of cervical disease, but the procedure requires controlled excision to obtain a complete cone-shaped specimen. This procedure can be technically demanding because CKC is affected by limited visibility, bleeding, anatomical variability, and the need to perform an angled excision while guiding the instrument accurately. A first version of the conisator was previously developed to support excision movements during CKC, but still required manual stabilisation of the bearing system. Therefore, the need for an integrated fixation mechanism formed the central design challenge addressed in this research.
A problem analysis was performed using literature research, input from gynaecologists, and evaluation of the first conisator prototype. Based on this analysis, a balloon-based fixation concept was selected and developed into a modular proof-of-concept prototype. The final design consisted of a manually inflated balloon retained within a bearing system, allowing temporary fixation around the cervix while the excision components could rotate. A phantom model was developed to simulate the cervicovaginal surgical field. The prototype was validated by eight gynaecologists through a simulated CKC procedure and questionnaire-based evaluation of task performance, design criteria, and requirements. In addition, a specimen measurement test was performed to assess the dimensional performance of the excised cone-shaped specimens.
The design criteria were generally evaluated positively, with an overall weighted score of 3.7 out of 5. Most procedural steps, including balloon inflation, stabiliser placement, and device removal, could be performed easily. The main design criteria requiring further optimisation were device control, cervix movement, cervical visibility, and user confidence. Cervical visibility through the inner ring remained limited, which made it more difficult to observe the
excision and locate the cervical canal during insertion. Overall, 11 of the 13 requirements were fulfilled. The two requirements that were not completely fulfilled concerned rotational resistance and anatomical fit. Rotational resistance received 63% positive responses, while anatomical fit was not fully achieved because the width of the fixation mechanism exceeded the reported mean minimum fornix width. The specimen measurement test showed that the conisator approximated the preset specimen width of 24 mm, with measured widths ranging from 21 to 26~mm and a mean width of 23.1 mm. Although minor variation remained between measurements, the specimen widths stayed close to the preset target. This indicates that the conisator effectively guided the excision towards the intended width, despite sensitivity to alignment, device orientation, blade sharpness, and scalpel-holder flexibility. From an LMIC perspective, the design fulfilled several key functions through electricity-independent and tool-free operation, modular construction, cost efficiency, design for sterilisation, and ease of use.
Overall, this research demonstrated that an integrated balloon-based fixation mechanism can provide temporary fixation and support controlled use of the conisator during simulated CKC. By integrating fixation into the instrument, this work addresses the design gap of stabilising the conisator at depth within the cervicovaginal space. The final proof-of-concept prototype therefore provides a promising basis for further development towards a reusable instrument for supporting CKC, particularly in LMIC settings. ...
A problem analysis was performed using literature research, input from gynaecologists, and evaluation of the first conisator prototype. Based on this analysis, a balloon-based fixation concept was selected and developed into a modular proof-of-concept prototype. The final design consisted of a manually inflated balloon retained within a bearing system, allowing temporary fixation around the cervix while the excision components could rotate. A phantom model was developed to simulate the cervicovaginal surgical field. The prototype was validated by eight gynaecologists through a simulated CKC procedure and questionnaire-based evaluation of task performance, design criteria, and requirements. In addition, a specimen measurement test was performed to assess the dimensional performance of the excised cone-shaped specimens.
The design criteria were generally evaluated positively, with an overall weighted score of 3.7 out of 5. Most procedural steps, including balloon inflation, stabiliser placement, and device removal, could be performed easily. The main design criteria requiring further optimisation were device control, cervix movement, cervical visibility, and user confidence. Cervical visibility through the inner ring remained limited, which made it more difficult to observe the
excision and locate the cervical canal during insertion. Overall, 11 of the 13 requirements were fulfilled. The two requirements that were not completely fulfilled concerned rotational resistance and anatomical fit. Rotational resistance received 63% positive responses, while anatomical fit was not fully achieved because the width of the fixation mechanism exceeded the reported mean minimum fornix width. The specimen measurement test showed that the conisator approximated the preset specimen width of 24 mm, with measured widths ranging from 21 to 26~mm and a mean width of 23.1 mm. Although minor variation remained between measurements, the specimen widths stayed close to the preset target. This indicates that the conisator effectively guided the excision towards the intended width, despite sensitivity to alignment, device orientation, blade sharpness, and scalpel-holder flexibility. From an LMIC perspective, the design fulfilled several key functions through electricity-independent and tool-free operation, modular construction, cost efficiency, design for sterilisation, and ease of use.
Overall, this research demonstrated that an integrated balloon-based fixation mechanism can provide temporary fixation and support controlled use of the conisator during simulated CKC. By integrating fixation into the instrument, this work addresses the design gap of stabilising the conisator at depth within the cervicovaginal space. The final proof-of-concept prototype therefore provides a promising basis for further development towards a reusable instrument for supporting CKC, particularly in LMIC settings. ...
Cervical cancer represents an important global health challenge, particularly in low- and middle-income countries (LMICs), where access to screening, treatment, and specialised equipment may be limited. Cold knife conisation (CKC) can play an important role in the treatment and diagnosis of cervical disease, but the procedure requires controlled excision to obtain a complete cone-shaped specimen. This procedure can be technically demanding because CKC is affected by limited visibility, bleeding, anatomical variability, and the need to perform an angled excision while guiding the instrument accurately. A first version of the conisator was previously developed to support excision movements during CKC, but still required manual stabilisation of the bearing system. Therefore, the need for an integrated fixation mechanism formed the central design challenge addressed in this research.
A problem analysis was performed using literature research, input from gynaecologists, and evaluation of the first conisator prototype. Based on this analysis, a balloon-based fixation concept was selected and developed into a modular proof-of-concept prototype. The final design consisted of a manually inflated balloon retained within a bearing system, allowing temporary fixation around the cervix while the excision components could rotate. A phantom model was developed to simulate the cervicovaginal surgical field. The prototype was validated by eight gynaecologists through a simulated CKC procedure and questionnaire-based evaluation of task performance, design criteria, and requirements. In addition, a specimen measurement test was performed to assess the dimensional performance of the excised cone-shaped specimens.
The design criteria were generally evaluated positively, with an overall weighted score of 3.7 out of 5. Most procedural steps, including balloon inflation, stabiliser placement, and device removal, could be performed easily. The main design criteria requiring further optimisation were device control, cervix movement, cervical visibility, and user confidence. Cervical visibility through the inner ring remained limited, which made it more difficult to observe the
excision and locate the cervical canal during insertion. Overall, 11 of the 13 requirements were fulfilled. The two requirements that were not completely fulfilled concerned rotational resistance and anatomical fit. Rotational resistance received 63% positive responses, while anatomical fit was not fully achieved because the width of the fixation mechanism exceeded the reported mean minimum fornix width. The specimen measurement test showed that the conisator approximated the preset specimen width of 24 mm, with measured widths ranging from 21 to 26~mm and a mean width of 23.1 mm. Although minor variation remained between measurements, the specimen widths stayed close to the preset target. This indicates that the conisator effectively guided the excision towards the intended width, despite sensitivity to alignment, device orientation, blade sharpness, and scalpel-holder flexibility. From an LMIC perspective, the design fulfilled several key functions through electricity-independent and tool-free operation, modular construction, cost efficiency, design for sterilisation, and ease of use.
Overall, this research demonstrated that an integrated balloon-based fixation mechanism can provide temporary fixation and support controlled use of the conisator during simulated CKC. By integrating fixation into the instrument, this work addresses the design gap of stabilising the conisator at depth within the cervicovaginal space. The final proof-of-concept prototype therefore provides a promising basis for further development towards a reusable instrument for supporting CKC, particularly in LMIC settings.
A problem analysis was performed using literature research, input from gynaecologists, and evaluation of the first conisator prototype. Based on this analysis, a balloon-based fixation concept was selected and developed into a modular proof-of-concept prototype. The final design consisted of a manually inflated balloon retained within a bearing system, allowing temporary fixation around the cervix while the excision components could rotate. A phantom model was developed to simulate the cervicovaginal surgical field. The prototype was validated by eight gynaecologists through a simulated CKC procedure and questionnaire-based evaluation of task performance, design criteria, and requirements. In addition, a specimen measurement test was performed to assess the dimensional performance of the excised cone-shaped specimens.
The design criteria were generally evaluated positively, with an overall weighted score of 3.7 out of 5. Most procedural steps, including balloon inflation, stabiliser placement, and device removal, could be performed easily. The main design criteria requiring further optimisation were device control, cervix movement, cervical visibility, and user confidence. Cervical visibility through the inner ring remained limited, which made it more difficult to observe the
excision and locate the cervical canal during insertion. Overall, 11 of the 13 requirements were fulfilled. The two requirements that were not completely fulfilled concerned rotational resistance and anatomical fit. Rotational resistance received 63% positive responses, while anatomical fit was not fully achieved because the width of the fixation mechanism exceeded the reported mean minimum fornix width. The specimen measurement test showed that the conisator approximated the preset specimen width of 24 mm, with measured widths ranging from 21 to 26~mm and a mean width of 23.1 mm. Although minor variation remained between measurements, the specimen widths stayed close to the preset target. This indicates that the conisator effectively guided the excision towards the intended width, despite sensitivity to alignment, device orientation, blade sharpness, and scalpel-holder flexibility. From an LMIC perspective, the design fulfilled several key functions through electricity-independent and tool-free operation, modular construction, cost efficiency, design for sterilisation, and ease of use.
Overall, this research demonstrated that an integrated balloon-based fixation mechanism can provide temporary fixation and support controlled use of the conisator during simulated CKC. By integrating fixation into the instrument, this work addresses the design gap of stabilising the conisator at depth within the cervicovaginal space. The final proof-of-concept prototype therefore provides a promising basis for further development towards a reusable instrument for supporting CKC, particularly in LMIC settings.