Design of a Flow-Powered Pulsating System for Equal Distribution in Multi-Nozzle Sterilization Rods

Master Thesis (2026)
Author(s)

B.J. van Neutegem (TU Delft - Mechanical Engineering)

Contributor(s)

T. Horeman – Mentor (TU Delft - Mechanical Engineering)

B. van Straten – Mentor (TU Delft - Mechanical Engineering)

J. Dankelman – Mentor (TU Delft - Mechanical Engineering)

Faculty
Mechanical Engineering
More Info
expand_more
Publication Year
2026
Language
English
Graduation Date
29-01-2026
Awarding Institution
Delft University of Technology
Programme
Mechanical Engineering, BioMechanical Design
Faculty
Mechanical Engineering
Page Views
111
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Distal nozzles in multioutlet sterilization rods receive reduced pressure due to axial pressure losses, limiting cleaning capacity when outlets are plugged to maintain performance. This thesis develops and evaluates a passive, flow-powered rotary-valve concept (GreenClean) intended to generate low-frequency pulsating jets while improving outlet-to-outlet distribution within a half-DIN sterilization basket. A five-outlet prototype was designed and experimentally validated using time-resolved pressure and flow measurements, progressive outlet-blockage tests, and supporting CFD as an idealised leakage-free baseline.

The prototype produced stable pulsation at approximately one cycle per 62–69 s (14–16 mHz), well below the 2 Hz requirement, and pulsation persisted under progressive and full blockage. However, uniform distribution was not achieved experimentally: mean deviations reached ∼ 25% in outlet flow and ∼ 39% in outlet pressure. In contrast, sealed CFD predicted outlet total-pressure deviations within ±4.1%, indicating that the concept can approach the distribution target in principle. Additional tests revealed substantial nonideal behaviour, including a leakage fraction of ηleak ≈ 0.48 at the connection/ sealing region and a marked reduction in available head in a jet height comparison. A stall-torque estimate of τstall ≈ 0.28 Nm provides an order-of-magnitude design margin for future sealing and alignment improvements. Environmental resistance was not validated and requires dedicated material and seal selection for washer-disinfector conditions.