We demonstrate the feasibility of intravascular ultrasound (IVUS) chirp imaging as well as chirp reversal ultrasound contrast imaging at intravascular ultrasound frequency. Chirp excitations were emitted with a 34 MHz single crystal intravascular transducer and compared to conventional Gaussian-shaped pulses of equal acoustic pressure. The signal to noise ratio of the chirp images was increased by up to 9 dB relative to the conventional images. Imaging of contrast microbubbles was implemented by chirp reversal, achieving a contrast to tissue ratio of 12 dB. The method shows potential for intravascular imaging of structures in and beyond coronary atherosclerotic plaques including vasa vasorum.
@enPhotoacoustic signals originating from weak sources can be hard to discriminate from higher intensity signals resulting from the photoacoustic background. In order to reveal these weaker signals we propose a method where the expected background signal is subtracted from the actual signals. In this method an ultrasound image provides the geometry for the pressure distribution used in a photoacoustic wave field simulation. The simulated photoacoustic signals are subtracted from the actual recordings and the residual is used for image reconstruction. This method was successfully validated experimentally with a vessel phantom containing three optical absorption irregularities within the vessel wall.