D.H. Plettenburg
Please Note
43 records found
1
Design of new hand prosthesis with two modes
Realizing a proof-of-principle to validate feasibility of a hybrid hand prosthesis
An upper limb amputee currently has two choices for a body-powered prosthesis, a Voluntary Opening (VO) or Voluntary Closing (VC) hand prosthesis. Which type is best, depends on the task and the individual. Currently, there are no good options for VO and VC in one design.
Objectives
To design a proof-of-principle of a hand prosthesis and to validate the feasibility of the prosthesis. The hand prosthesis must have two different modes, the VO mode and the VC mode. Changing between these modes can be done without using the other hand..
Methods
First, a list of requirements and wishes was made, where cable forces are important because too much cable force can cause discomfort, fatigue or make the prosthesis difficult to control.
A new design was created and a prototype was built for user testing. SHAP was used as the outcome measure.
Results
The prototype allowed the user to change between VO and VC and SHAP showed promising Linear Index of Functionality (LIF) values. The calculated cable forces were mostly within the requirements. However, because the prototype of this design was produced with an FDM-printer, not all the tests could be done and the calculated forces could not be verified.
Conclusions
A prototype of a hybrid hand prosthesis with VO and VC modes showed promise, with positive test results. However, limitations in the prototype's construction hindered some tests. The calculated cable forces mostly met requirements, except for a slightly high pinch grip force. Further development is needed, and there are material and design recommendations. This concept demonstrates feasibility but requires more work to become a functional hand prosthesis.
...
An upper limb amputee currently has two choices for a body-powered prosthesis, a Voluntary Opening (VO) or Voluntary Closing (VC) hand prosthesis. Which type is best, depends on the task and the individual. Currently, there are no good options for VO and VC in one design.
Objectives
To design a proof-of-principle of a hand prosthesis and to validate the feasibility of the prosthesis. The hand prosthesis must have two different modes, the VO mode and the VC mode. Changing between these modes can be done without using the other hand..
Methods
First, a list of requirements and wishes was made, where cable forces are important because too much cable force can cause discomfort, fatigue or make the prosthesis difficult to control.
A new design was created and a prototype was built for user testing. SHAP was used as the outcome measure.
Results
The prototype allowed the user to change between VO and VC and SHAP showed promising Linear Index of Functionality (LIF) values. The calculated cable forces were mostly within the requirements. However, because the prototype of this design was produced with an FDM-printer, not all the tests could be done and the calculated forces could not be verified.
Conclusions
A prototype of a hybrid hand prosthesis with VO and VC modes showed promise, with positive test results. However, limitations in the prototype's construction hindered some tests. The calculated cable forces mostly met requirements, except for a slightly high pinch grip force. Further development is needed, and there are material and design recommendations. This concept demonstrates feasibility but requires more work to become a functional hand prosthesis.
Haptic feedback in upper limb prostheses
Development and evaluation of a pneumatic haptic device providing feedback for upper limb prostheses
An active thumb module for the Symbihand
Assisting Duchenne Muscular Dystrophy patients in ADL
A Dual Actuated Hand Exoskeleton
Improving Force Transmission
Its complicating that on the hand little space is available to build an exoskeleton. So, moment arms of actuators must be small. This complicates generating adequate moments. Strong actuators may be a solution, but they add weight, which should be minimized for comfort. The goal of this thesis is to design a proof of concept that improves force output, considering weight and size criteria.
The proof of concept divides the function of the exoskeleton into two phases that work one after the other. In the first phase the fingers flex to encapsulate an object, in the second phase force is applied. Each phase has an actuator that suits its requirements.
Results show that force output equals 5.087 [N], while the system weights 49.9 [g]. This equals a force-to-weight ratio of 102 [N/kg].
Force output was lower than theorized due to a variety of reasons and did not meet the set criteria of 10 [N]. The criteria of size were not met at the dorsal side of the proximal phalanx. The criteria were 20 [mm] extrusion, 27 [mm] was the end result. Both failures are expected to be solvable. It is concluded that with updates, the design can improve force output while maintaining a small and lightweight design. ...
Its complicating that on the hand little space is available to build an exoskeleton. So, moment arms of actuators must be small. This complicates generating adequate moments. Strong actuators may be a solution, but they add weight, which should be minimized for comfort. The goal of this thesis is to design a proof of concept that improves force output, considering weight and size criteria.
The proof of concept divides the function of the exoskeleton into two phases that work one after the other. In the first phase the fingers flex to encapsulate an object, in the second phase force is applied. Each phase has an actuator that suits its requirements.
Results show that force output equals 5.087 [N], while the system weights 49.9 [g]. This equals a force-to-weight ratio of 102 [N/kg].
Force output was lower than theorized due to a variety of reasons and did not meet the set criteria of 10 [N]. The criteria of size were not met at the dorsal side of the proximal phalanx. The criteria were 20 [mm] extrusion, 27 [mm] was the end result. Both failures are expected to be solvable. It is concluded that with updates, the design can improve force output while maintaining a small and lightweight design.
A New Pneumatic Actuator Design
For Compliant and Lightweight Lower Limb Exoskeletons
the performance, pneumatic actuators can be used as an alternative to the commonly used electromechanical actuators. This thesis presents a new pneumatic actuator design to power a lower limb exoskeleton in daily life
activities. The goal of this thesis is to take a step towards the design and development of exoskeletons with reduced weight, increased compliance and that enable paraplegic patients to perform daily life activities. Out of several
concepts, a single vane rotary actuator is worked out in more detail and a prototype is manufactured. This actuator is able to generate a peak torque of 96 ± 2 [Nm] at an operating pressure of 12 [bar], has a range of motion of
132 ± 1 [°] and is able to generate a peak angular velocity of 430 ± 5 [°/s] and a peak power of 252 ± 3 [W]. The designed actuator has a width of 165 [mm], a length of 135 [mm], a dept of 54 [mm] and a total mass of 2.3 [kg], but
with small changes can be decreased to below 1.6 [kg]. This is the first known pneumatic actuator able to generate sufficient dynamics to power a full mobilization lower limb exoskeleton in daily life activities. It can be concluded
that this new design is a promising actuator to be used in lower limb exoskeletons. In further work, the bandwidth of the actuator must be determined accurately. ...
the performance, pneumatic actuators can be used as an alternative to the commonly used electromechanical actuators. This thesis presents a new pneumatic actuator design to power a lower limb exoskeleton in daily life
activities. The goal of this thesis is to take a step towards the design and development of exoskeletons with reduced weight, increased compliance and that enable paraplegic patients to perform daily life activities. Out of several
concepts, a single vane rotary actuator is worked out in more detail and a prototype is manufactured. This actuator is able to generate a peak torque of 96 ± 2 [Nm] at an operating pressure of 12 [bar], has a range of motion of
132 ± 1 [°] and is able to generate a peak angular velocity of 430 ± 5 [°/s] and a peak power of 252 ± 3 [W]. The designed actuator has a width of 165 [mm], a length of 135 [mm], a dept of 54 [mm] and a total mass of 2.3 [kg], but
with small changes can be decreased to below 1.6 [kg]. This is the first known pneumatic actuator able to generate sufficient dynamics to power a full mobilization lower limb exoskeleton in daily life activities. It can be concluded
that this new design is a promising actuator to be used in lower limb exoskeletons. In further work, the bandwidth of the actuator must be determined accurately.
Phase switching dynamics for body-powered hand prostheses
Design of a booster to passively switch between phase dynamics
results show that there is room for improvement regarding the cable operating force and locking mechanism. At larger apertures the cable operating force necessary to get an adequate grip force slightly exceeds comfortable limits. The VC WILMER appealing prehensor is equipped with the vertical mount Sure-Lok.
Although the vertical mount Sure-Lock is a more effective locking mechanism than locking mechanisms tested in literature, the drop in grip force induced after activation is still quite large. ...
results show that there is room for improvement regarding the cable operating force and locking mechanism. At larger apertures the cable operating force necessary to get an adequate grip force slightly exceeds comfortable limits. The VC WILMER appealing prehensor is equipped with the vertical mount Sure-Lok.
Although the vertical mount Sure-Lock is a more effective locking mechanism than locking mechanisms tested in literature, the drop in grip force induced after activation is still quite large.
Therefore, the aim of this paper is to design a breathable, volume adjustable, parametric, 3D printed socket for transradial amputees in low-income countries.
This new design consists of straps connected with hinges which improve the range of motion. The material used for the socket is TPU 95A. The straps are perforated to improve the breathability of the socket. Furthermore, a clickable spacer, which is strong enough to withstand a force of 200 N, is attached to make the socket volume adjustable. The complete socket is also subjected to strength tests which resulted in a maximum manageable force of 100 N, attached to the prosthetic hand. However, the socket has not been evaluated by patients, which means that real-life wearing conditions are not achieved. Therefore, it would be a recommendation for future research to examine the socket on patients. Overall, this study provides the design of an affordable, comfortable, aesthetically pleasing, durable, parametric socket for amputees in low-income countries. ...
Therefore, the aim of this paper is to design a breathable, volume adjustable, parametric, 3D printed socket for transradial amputees in low-income countries.
This new design consists of straps connected with hinges which improve the range of motion. The material used for the socket is TPU 95A. The straps are perforated to improve the breathability of the socket. Furthermore, a clickable spacer, which is strong enough to withstand a force of 200 N, is attached to make the socket volume adjustable. The complete socket is also subjected to strength tests which resulted in a maximum manageable force of 100 N, attached to the prosthetic hand. However, the socket has not been evaluated by patients, which means that real-life wearing conditions are not achieved. Therefore, it would be a recommendation for future research to examine the socket on patients. Overall, this study provides the design of an affordable, comfortable, aesthetically pleasing, durable, parametric socket for amputees in low-income countries.
Design of a Pneumatically Powered Hand Prosthesis for Toddlers
Introducing an underactuated 3-DOF linkage-based finger transmission mechanism
Commonly, the main reasons for abandonment are related to comfort, control and cosmetics.
Prostheses are often experienced as too heavy and have limited functionality.
A pneumatic power source is able to provide a relatively high force using a small and lightweight actuator.
However, state-of-the-art pneumatically powered upper extremity prosthesis offer either high grasp forces or the ability to adapt to the size and shape of an object.
This article presents the design of a pneumatically powered hand prosthesis for toddlers, which focuses on being lightweight and highly functional.
An underactuated 3-DOF linkage-based finger transmission mechanism was created and served as a proof-of-principle.
The resulting prototype was 3D printed and tested following an elaborate list of criteria.
The results demonstrate that the mechanism is capable of transmitting an actuator force of 100.5 N to a fingertip force of 32.3 N using 5 bar of pressure.
The proposed mechanism also offers the functionality of adaptive grasping.
The conceptual design and prototype show promising capabilities for the development of a lightweight and highly functional prosthesis for toddlers. ...
Commonly, the main reasons for abandonment are related to comfort, control and cosmetics.
Prostheses are often experienced as too heavy and have limited functionality.
A pneumatic power source is able to provide a relatively high force using a small and lightweight actuator.
However, state-of-the-art pneumatically powered upper extremity prosthesis offer either high grasp forces or the ability to adapt to the size and shape of an object.
This article presents the design of a pneumatically powered hand prosthesis for toddlers, which focuses on being lightweight and highly functional.
An underactuated 3-DOF linkage-based finger transmission mechanism was created and served as a proof-of-principle.
The resulting prototype was 3D printed and tested following an elaborate list of criteria.
The results demonstrate that the mechanism is capable of transmitting an actuator force of 100.5 N to a fingertip force of 32.3 N using 5 bar of pressure.
The proposed mechanism also offers the functionality of adaptive grasping.
The conceptual design and prototype show promising capabilities for the development of a lightweight and highly functional prosthesis for toddlers.
The Hydraulic WILMER Tweezer
Design and Development of a Hydraulic Voluntary Closing Mechanism within a Body Powered Split-Hook Prosthesis
The Power of Pneumatics
Design of a Novel Elbow Prosthesis
Introduction: Almost 57 percent of above elbow amputees stop using their prosthesis. One of the main factors of non-wear is the high weight. Pneumatic actuation can be more lightweight than electronic actuation, and might therefore help with reducing the mass of a prosthesis and increase its comfort. To illustrate this potential, the goal of this thesis was: "To show the potential of pneumatic prostheses by designing a lightweight pneumatic elbow prosthesis, with a functionality equivalent to existing prostheses". Methods: The design process was divided into five phases: Analysis, Conceptualization, Embodiment, Manufacturing and Assembly, and Testing. The analysis brought to light the functions, requirements, wishes, and design values. The two design phases, conceptualization and embodiment, translated these into a potentially viable prosthesis. Afterwards, a prototype was manufactured, assembled, and tested to see if the functions, wishes, and requirements were met. Results: The prototype weighs almost 1300 [g]. A payload of 4.0 [kg] can be lifted to 87 [°]. In theory, 2.5 [kg] can be lifted throughout the entire range of motion, up to 140 [°]. The maximum pronation/supination torque is over 2.8 [Nm] throughout the entire range of motion of ± 90 [°]. The locking mechanism is theoretically capable of passively holding over 6 [kg] in any position. In theory, an average of 125 cycles can be achieved per 25 [g] CO2 cartridge. Discussion: The prototype does not have an integrated fuel source. The weight of the prototype exceeds the maximum by almost 280 [g]. Switching to electronic control significantly decreases the prosthesis mass. The frame of the prototype was under-constrained, leading to an increased friction for higher angles of flexion. Conclusion: The combination of the high functionality and low mass of the prototype shows potential for pneumatic actuation for prostheses. ...
Introduction: Almost 57 percent of above elbow amputees stop using their prosthesis. One of the main factors of non-wear is the high weight. Pneumatic actuation can be more lightweight than electronic actuation, and might therefore help with reducing the mass of a prosthesis and increase its comfort. To illustrate this potential, the goal of this thesis was: "To show the potential of pneumatic prostheses by designing a lightweight pneumatic elbow prosthesis, with a functionality equivalent to existing prostheses". Methods: The design process was divided into five phases: Analysis, Conceptualization, Embodiment, Manufacturing and Assembly, and Testing. The analysis brought to light the functions, requirements, wishes, and design values. The two design phases, conceptualization and embodiment, translated these into a potentially viable prosthesis. Afterwards, a prototype was manufactured, assembled, and tested to see if the functions, wishes, and requirements were met. Results: The prototype weighs almost 1300 [g]. A payload of 4.0 [kg] can be lifted to 87 [°]. In theory, 2.5 [kg] can be lifted throughout the entire range of motion, up to 140 [°]. The maximum pronation/supination torque is over 2.8 [Nm] throughout the entire range of motion of ± 90 [°]. The locking mechanism is theoretically capable of passively holding over 6 [kg] in any position. In theory, an average of 125 cycles can be achieved per 25 [g] CO2 cartridge. Discussion: The prototype does not have an integrated fuel source. The weight of the prototype exceeds the maximum by almost 280 [g]. Switching to electronic control significantly decreases the prosthesis mass. The frame of the prototype was under-constrained, leading to an increased friction for higher angles of flexion. Conclusion: The combination of the high functionality and low mass of the prototype shows potential for pneumatic actuation for prostheses.
ARM3D
A 3D printed transhumeral arm prosthesis
Sensitivity and control of a pneumatic force transducer
A proof of principle
In Chapter 2, the focus is on balancers applied in exoskeletons and robotic manipulators. The goal of this chapter is twofold. The first goal is to introduce the optimal balancer adjustment types for application in exoskeletons and robotic manipulators. The second goal is to compare their performance based on the balancing quality, energy involved in adjustment and dimensions through an analytical MATLAB model. The introduced adjustment types incorporate a commercial available spring with realistic properties and their performance is considered in their complete range of motion.
The results of the complete thesis research are discussed in Chapter 3, and in Chapter 4, the main conclusions are highlighted. This thesis ends with additional information about the performed research, which can be found in Appendices A to H. ...
In Chapter 2, the focus is on balancers applied in exoskeletons and robotic manipulators. The goal of this chapter is twofold. The first goal is to introduce the optimal balancer adjustment types for application in exoskeletons and robotic manipulators. The second goal is to compare their performance based on the balancing quality, energy involved in adjustment and dimensions through an analytical MATLAB model. The introduced adjustment types incorporate a commercial available spring with realistic properties and their performance is considered in their complete range of motion.
The results of the complete thesis research are discussed in Chapter 3, and in Chapter 4, the main conclusions are highlighted. This thesis ends with additional information about the performed research, which can be found in Appendices A to H.
The goal of this study is to improve the grasping performance of the VO/VC device. Changing fingers did not create a device which could fulfil all the requirements. Instead, another idea was embodied, prototyped and tested.
This device is able switch between VO and VC, by the use of two separate pivot points. Switching between the two is done with a bi-stable mechanism, which is activated by the intact hand. Results show that the device can be used in both modes and that users are able to switch with the bi-stable mechanism, but the mechanical properties do not meet the requirements. Additionally, the device performs less than expected on the user test, due to the bad mechanical properties and misalignments in the 3D-printed prototype. The device created is a good proof of principle for using two pivot points to shift between VO and VC. However, the bi-stable mechanism did not function as expected and a lot of friction is present leading to high forces in the cable. It is recommended to change the material, focus on using other springs which allow for larger opening widths and generate less friction and change the switching system to increase usability and safety. ...
The goal of this study is to improve the grasping performance of the VO/VC device. Changing fingers did not create a device which could fulfil all the requirements. Instead, another idea was embodied, prototyped and tested.
This device is able switch between VO and VC, by the use of two separate pivot points. Switching between the two is done with a bi-stable mechanism, which is activated by the intact hand. Results show that the device can be used in both modes and that users are able to switch with the bi-stable mechanism, but the mechanical properties do not meet the requirements. Additionally, the device performs less than expected on the user test, due to the bad mechanical properties and misalignments in the 3D-printed prototype. The device created is a good proof of principle for using two pivot points to shift between VO and VC. However, the bi-stable mechanism did not function as expected and a lot of friction is present leading to high forces in the cable. It is recommended to change the material, focus on using other springs which allow for larger opening widths and generate less friction and change the switching system to increase usability and safety.